You are here: Home » Products » Water Quality Instruments » Turbidity Meter » TP100 Digital Portable Turbidity Meter with Scattered Light

Share to:

TP100 Digital Portable Turbidity Meter with Scattered Light

Portable turbidity meter, up to 5 points calibration, selectable turbidity unit and measurement mode.

  • TP100
  • Dshing
  • Range: 0~1100 NTU
  • Accuracy: ±2% (0~500 NTU)
  • Resolution: 0.01 (0~100 NTU)
  • Stray Light: < 0.02 NTU
  • Display: LCD
  • Dimension: 180(L)*85(W)*70(H)mm
Quantity:

Product Description

TP100 Digital Portable Turbidity Meter with Scattered Light

Product Description
Portable turbidity meter, up to 5 points calibration, selectable turbidity unit and measurement mode.
TP100 Portable Turbidity Meter 1
Features

• Easy-to-use portable turbidity meter comes with a standard carrying case.
• 2 to 5 points push-button calibration using the Formazin Standards.
• Selectable 4 turbidity units, including the NTU, FNU, EBC and ASBC.
• Single measurement mode automatically recognizes and locks the stable reading.
• Continuous measurement mode can be used for indexing or matching the sample vials.
• Auto-Power Off effectively conserves battery life.
TP100 Portable Turbidity Meter 4

• Setup menu allows setting the 7 parameters, including the number of calibration points,
  resolution, date and time, etc.
• Reset feature automatically resumes all settings back to factory default options.
• Expanded memory stores and recalls up to 100 data sets.
• Stored data can be transferred into the computer by USB communication interface.
• Multi-mode power scheme (batteries, power adapter or USB port) ensures that use the
  meter smoothly.

Applications

• Drinking water and swimming pool
• Environmental monitoring

Specifications

SN. Model TB100
1 Measurement Method ISO 7027 nephelometric method (90° scattered light)
2 Range 0~1100 NTU, 0~275 EBC, 0~9999 ASBC
3 Accuracy ±2% (0~500 NTU), ±3% (501~1100 NTU)
4 Resolution 0.01 (0~100 NTU), 0.1 (100~999 NTU), 1 (999~1100 NTU)
5 Calibration Points 2 to 5 points (0.02, 10, 200, 500, 1000 NTU)
6 Light Source Infrared-emitting diode, 850nm wavelength
7 Detector Silicon photodiode
8 Stray Light < 0.02 NTU
9 Suitable Vials Round cuvettes, 60×25mm
10 Power Off Manual or Automatic (15 minutes after last key pressed)
11 Memory Stores up to 100 data sets
12 Output USB communication interface
13 Display LCD
14 Power Off 3×1.5V AA batteries or DC5V power adapter
15 Dimensions 180(L)×85(W)×70(H)mm
16 Weight 300g

Other Information
Order Code Accessories
TB100 Turbidity standards (0, 200, 500, 1000 NTU)
Sample vials (2 PCS)
Lint-free cloth
Carrying case

TP100 Portable Turbidity Meter 3
WhatsApp:+86-13286053265
Tel: 86-756-8305811  Email: admin@dshing.com  