DSHD-261 Manual Pensky-Martens Closed Cup Flash Point Tester ASTM D93
Product Description
DSHD-261 Manual Pensky-Martens Closed Cup Flash Point Tester ASTM D93
Summary
The DSHD-261 Manual Pensky-Martens Closed Cup Flash Point Tester is designed and manufactured according to GB/T 261-2021 and ASTM D93. It determines the closed cup flash point of petroleum products. As a manual model, heating power is continuously adjustable from 0 to 600W and shown visually by voltmeter, with heating mode adjusted manually. Stirring is mechanical drive, meeting Procedure A/C (90–120 rpm) and Procedure B (250±10 rpm) requirements. The silicon carbide furnace body and stainless-steel table board make the structure small, exquisite, and easy to clean. It comes with mercury-in-glass thermometers in three ranges: -5–110°C (0.5°C division), 20–150°C (1°C division), and 90–370°C (2°C division). The oil cup has an inner diameter of 50.7–50.8mm, depth of 55.7–56.0mm, and approximately 70ml capacity. Ignition source is gas or other civilian fuels. This manual model is the most economical choice, suitable for low-volume or occasional testing where full automation is not required. For higher throughput, semi-automatic model DSHD-261-1 and fully automatic model DSHD-261D are also available.
Technical Data Sheet
I. Main technical features
1. This instrument is designed as per GB/T261 and ASTM D93.
2. The heating power is continuously adjustable. The power is shown visually by voltmeter.The temperature control mode is advanced and reasonable.
3. The structure is designed small and exquisite.The stainless-steel table board is beautiful and easy to clean.
4. Operation is easy.Test results are precise.It has been chosen as the instrument to test the closed cup flash point by many metering and detecting institutes.
| 1 | Power supply: | AC 220V±10%, 50 Hz; |
| 2 | Heating device | |
| 2.1 | Furnace power is; itsis | 600 W, made of SiC materials |
| 2.2 | Heating power adjustable range | 0 W~600W |
| 3 | Sample temperature rising rate | 1℃~12℃ per minute, manually controlled |
| 4 | Electric stirring device | |
| 4.1 | Stirring motor | 45TYZ Motor |
| 4.2 | Driving device | By a soft shaft |
| 4.3 | Stirring blade specification | 8 mm×40 mm |
| 5 | Stirring speed | As per the requirement of GB/T261-2008 |
| 5.1 | Procedure A; | 90 RPM~120 RPM |
| 5.2 | Procedure B; | 250±10 RPM |
| 6 | Standard oil cup | As per the requirement of GB/T261-2008 |
| 6.1 | Inner diameter | 50.7 mm~50.8 mm |
| 6.2 | Depth | 55.7 mm~56.0 mm |
| 6.3 | Depth of marked line in the oil cup | 33.9 mm~34.3 mm |
| 6.4 | Oil sample cubage | About 70 ml |
| 7 | Flame application device | |
| 7.1 | Gas source | Coal gas (or other civilian combustible gas) |
| 7.2 | Aperture of flame application device | 0.8 mm |
| 8 | Thermometer | Rod type Mercury Thermometer, as per GB/T261-2008 |
| 8.1 | Temperature range | -5℃~110℃, Each scale division is 0.5 ℃ |
| 8.2 | Temperature range | Temperature range: 20℃~150℃, Each scale division is 1.0 ℃ |
| 8.3 | Temperature range | 90℃~370℃, Each scale division is 2 ℃ |
| 9 | Ambient temperature | ≤35℃ |
| 10 | Relative humidity | ≤85% |
| 11 | Total power consumption | Not more than 650 W |