DSHD-3536-1 Semi-Automatic Cleveland Open-Cup Flash Point Tester

Product Description

Summary
This instrument is designed and made as per GB/T3536-2008 Petroleum Products - Determination of Flash and Fire Points - Cleveland open cup method and ASTM D 92 Standard Test Method for Flash and Fire Points by Cleveland Open Cup Tester. It is applicable to all petroleum products with flash points above 79℃ and below 400℃ except fuel oils.
 

Technical Data Sheet

I. Main technical features
1. It adopts technology of single chip microcomputer and LCD screen. English man-machine interface.The heating device is continuously adjustable.Easy to set parameters.
2. The LCD screen has prompt menu, prompt type input for operation interface. It shows set parameters and real-time display sample temperature and other parameters. Press the record key when flash point appearing. The screen will display and save flash point value.
3. It is newly design and small structure. Stainless steel working table make it elegant and beautiful. The operator can do determination as long as connecting with coal gas or other civil gas. Easy to use and results are exact.
4. It is equipped with wind-shelter and flame extinguishing cover which are accord with requirements of test.
Accurate heating rate.The instrument can do test automatically.
5. Operator only need to observe the flash point appearing. The cost performance is high.

II. Main technical specifications

SN. Item Parameter
1  Power supply  Power supply:  AC 220 V±10%, 50 Hz
2  Heating device  Electric furnace. No naked fire. Anti- explosion.The power can be continuously 
adjusted from 0 to 600W.The  maximum heating temperature is 400℃.
3  Temperature control  SCM temperature control. The heating rate  can meet standard GB/T 3536-2008  
4  Temperature display  LCD screen shows kinds of parameters. The showing range is (0~400)℃,precision 0.1℃
5  Flame application device  Automatic
6  Temperature sensor  RTD, Pt100
7  Ignition device
7.1  Ignition source gas or other civil gas
7.2  Aperture of spout About 0.8 mm
8  Ambient temperature  -10 ℃~50 ℃
9  Relative humidity   ≤85%
10  Max power consumption  650W
11  Overall dimension Overall dimension: 340mm×320mm×450mm(Temperature sensor is included.)

Product Inquiry

DSHD-5208D Rapid Low Temperature Closed Cup Flash Point Tester

DSHD-0633 Liquid Petroleum Asphalt Flash Point Tester (Tag Open Cup Method)

DSHD-5208A High-temperature Flash Point Tester

DSHD-5208 Rapid Closed Cup Flash Point Tester

DSHD-261 Pensky-Martens Closed Cup Flash Point Tester