Quantity: | |
---|---|
Product Description
Technical Data Sheet
I. Main technical features
1.Designed as per GB/T 5208-2008 Determination of flash point - Rapid equilibrium closed cup method. It can realize purpose of testing closed cup flash point rapidly and at low temperature.
2.The instrument adopts semiconductor refrigeration device and external water cooling device.The size is small and cooling rate
is rapid.Easy to operate.
3.Test sample needed is less.2ml for each time(4ml for solid sample and semi-solid sample)
4.The determination procedure is automatically completed except ignition.Test results can be printed automatically.
II. Main technical specifications and parameters
SN. | Item | Parameter |
1 | Power supply | AC(220±10%)V, 50Hz |
2 | Flash point measuring range | 50℃~100℃ |
3 | Precision of determination | |
3.1 | Absolute difference(same operator) | Llower than 2℃(same operator) |
3.2 | Absolute difference(different operator) | Lower than 3℃(different operator) |
4 | Temperature control precision | ±0.5℃ |
5 | Igniting device | Electric igniting gun |
6 | Cooling mode | Semiconductor(with cold water cycle at external connection) |
7 | Maximum power consumption | 300W |
8 | Ambient temperature | 5℃~30℃ |
9 | Relative humidity | 30%~80% |
10 | Dimension | 370mm×280mm×280mm |