Product Description
Technical Data Sheet
I. Introduction
The product is deigned per as the basic principles of the emission spectrum , widely used in agricultural fertilizers, soil analysis, concrete analysis, industry analysis and determination silicic acid, health care clinical testing and pathology.
II.Product Features
1. K, Na can be tested simultaneously , it also can test the lithium element
2. LCD display, digital direct reading,
3. With optional size of the flame function and automatic flame-out protection device .
4. Simple, rapid analysis, the sample is less
5. Gas:Liquefied petroleum gas
III.The main technical indicators
1. Stability:
Consecutive injections of standard solutions, the biggest change in the relative volume in 15S ≤ 3%
Test once per minute, test six times in total, the biggest change in the relative volume ≤ 15%
2. Repeatability:
Conduct 7 consecutive independent measurements for one standard solution,the relative standard deviation: ≤ 3%
3. Linearity error:
K: ≤ 0.005mmol / L (0.0100 ~ 0.0800) mmol / L
Na: ≤ 0.03mmol / L (0.0500 ~ 0.400) mmol / L
4. Detection limit:
K: The detection limit ≤ 0.004mmol / L; Na: detection limit ≤ 0.008mmol / L
5. Response time: Response time <8S
6. Suction jet of samples: Suction jet of samples <6mL/min
Other:
1. Complete Set: one host, one air compressor.
2. Normal working conditions:
Ambient temperature: 10 ℃ ~ 35 ℃
Relative humidity: ≤ 85%
The instrument should be placed horizontally on table with no vibration, avoid direct exposuring to bright light, keep away from strong electromagnetic interference, no strong airstream.
Power supply: AC220 ± 22V, Frequency: 50HZ ± 1 HZ, with good ground connection.
Rated power: 250W